P2.04 ASSESSMENT OF BRACHIAL ARTERY REACTIVITY USING THE ENDOCHECK: REPEATABILITY, REPRODUCIBILITY AND PRELIMINARY COMPARISON WITH ULTRASOUND
- DOI
- 10.1016/j.artres.2013.10.066How to use a DOI?
- Abstract
Objective: The Endocheck, embedded within the Vicorder device, uses cuff-based, pulse volume (PV) displacement to record brachial PV waveforms at baseline, and during reactive hyperaemia. The aim of this study was to assess the utility of the Endocheck method.
Methods: The study consisted of two parts. Part 1: Healthy volunteers (n=9) were studied twice, separated by 24hours. Each visit consisted of two studies, 30min apart, where, after 10min supine rest, brachial BP was assessed and PV waveforms recorded for 10sec (baseline). A cuff placed distally around the forearm was then inflated to 200mmHg for 5min. Following cuff-release, PV waveforms were recorded for 3min. The square root of the ratio of peak:baseline PV during hyperaemia (√V2/V1) was calculated. Part 2: Healthy volunteers (n=16) were studied once. Brachial artery responses were assessed simultaneously in both arms, using ultrasound (right arm) and Endocheck (left arm), following a similar protocol as above.
Results: Part 1: An average change in PV of 74±82% was detected in response to forearm ischaemia (P=0.003). Within-visit repeatability was acceptable, with a mean (±SD) difference in √V2/V1 of 0.03±0.25 (P=0.6), and a high correlation between studies (r=0.64; P=0.004). Between-visit reproducibility was high, with a mean difference of 0.004±0.17 (P=0.9) and a strong correlation between readings (r=0.81; P<0.0001; Figure 1). Part 2: There was a modest association (r=0.14, P=NS) between hyperaemic responses assessed using the different methods.
Conclusion: The Endocheck can reliably assess changes in brachial PV during hyperaemia. Further studies are required to determine whether the observed changes reflect endothelial function.
- Open Access
- This is an open access article distributed under the CC BY-NC license.
Download article (PDF)
View full text (HTML)
Cite this article
TY - JOUR AU - L.M. Day AU - K.M. Maki-Petaja AU - I.B. Wilkinson AU - C.M. McEniery PY - 2013 DA - 2013/11/11 TI - P2.04 ASSESSMENT OF BRACHIAL ARTERY REACTIVITY USING THE ENDOCHECK: REPEATABILITY, REPRODUCIBILITY AND PRELIMINARY COMPARISON WITH ULTRASOUND JO - Artery Research SP - 119 EP - 120 VL - 7 IS - 3-4 SN - 1876-4401 UR - https://doi.org/10.1016/j.artres.2013.10.066 DO - 10.1016/j.artres.2013.10.066 ID - Day2013 ER -