Artery Research

Volume 7, Issue 3-4, September 2013, Pages 119 - 120

P2.04 ASSESSMENT OF BRACHIAL ARTERY REACTIVITY USING THE ENDOCHECK: REPEATABILITY, REPRODUCIBILITY AND PRELIMINARY COMPARISON WITH ULTRASOUND

Authors
L.M. Day, K.M. Maki-Petaja, I.B. Wilkinson, C.M. McEniery
University of Cambridge, Cambridge, United Kingdom
Available Online 11 November 2013.
DOI
10.1016/j.artres.2013.10.066How to use a DOI?
Abstract

Objective: The Endocheck, embedded within the Vicorder device, uses cuff-based, pulse volume (PV) displacement to record brachial PV waveforms at baseline, and during reactive hyperaemia. The aim of this study was to assess the utility of the Endocheck method.

Methods: The study consisted of two parts. Part 1: Healthy volunteers (n=9) were studied twice, separated by 24hours. Each visit consisted of two studies, 30min apart, where, after 10min supine rest, brachial BP was assessed and PV waveforms recorded for 10sec (baseline). A cuff placed distally around the forearm was then inflated to 200mmHg for 5min. Following cuff-release, PV waveforms were recorded for 3min. The square root of the ratio of peak:baseline PV during hyperaemia (√V2/V1) was calculated. Part 2: Healthy volunteers (n=16) were studied once. Brachial artery responses were assessed simultaneously in both arms, using ultrasound (right arm) and Endocheck (left arm), following a similar protocol as above.

Results: Part 1: An average change in PV of 74±82% was detected in response to forearm ischaemia (P=0.003). Within-visit repeatability was acceptable, with a mean (±SD) difference in √V2/V1 of 0.03±0.25 (P=0.6), and a high correlation between studies (r=0.64; P=0.004). Between-visit reproducibility was high, with a mean difference of 0.004±0.17 (P=0.9) and a strong correlation between readings (r=0.81; P<0.0001; Figure 1). Part 2: There was a modest association (r=0.14, P=NS) between hyperaemic responses assessed using the different methods.

Conclusion: The Endocheck can reliably assess changes in brachial PV during hyperaemia. Further studies are required to determine whether the observed changes reflect endothelial function.

Open Access
This is an open access article distributed under the CC BY-NC license.

Journal
Artery Research
Volume-Issue
7 - 3-4
Pages
119 - 120
Publication Date
2013/11/11
ISSN (Online)
1876-4401
ISSN (Print)
1872-9312
DOI
10.1016/j.artres.2013.10.066How to use a DOI?
Open Access
This is an open access article distributed under the CC BY-NC license.

Cite this article

TY  - JOUR
AU  - L.M. Day
AU  - K.M. Maki-Petaja
AU  - I.B. Wilkinson
AU  - C.M. McEniery
PY  - 2013
DA  - 2013/11/11
TI  - P2.04 ASSESSMENT OF BRACHIAL ARTERY REACTIVITY USING THE ENDOCHECK: REPEATABILITY, REPRODUCIBILITY AND PRELIMINARY COMPARISON WITH ULTRASOUND
JO  - Artery Research
SP  - 119
EP  - 120
VL  - 7
IS  - 3-4
SN  - 1876-4401
UR  - https://doi.org/10.1016/j.artres.2013.10.066
DO  - 10.1016/j.artres.2013.10.066
ID  - Day2013
ER  -