Haemolysis in Centrifugal Pump of Artificial Blood Circulation System
Authors
A.V. Bobkov
Corresponding Author
A.V. Bobkov
Available Online 3 October 2020.
- DOI
- 10.2991/ahsr.k.201001.020How to use a DOI?
- Keywords
- artificial blood circulation system, centrifugal pump, haemolysis
- Abstract
The article deals with possible causes of increased haemolysis in the centrifugal pump of the artificial blood circulation systems. One of the hydrodynamic reasons is indicated – a strong vortex formation of the blood flow inside the pump impeller, which occurs when the impeller rotates. This factor is a natural demonstration of the mechanism of converting the mechanical energy of the impeller rotation into the pressure of the pumped blood flow.
- Copyright
- © 2020, the Authors. Published by Atlantis Press.
- Open Access
- This is an open access article distributed under the CC BY-NC license (http://creativecommons.org/licenses/by-nc/4.0/).
Cite this article
TY - CONF AU - A.V. Bobkov PY - 2020 DA - 2020/10/03 TI - Haemolysis in Centrifugal Pump of Artificial Blood Circulation System BT - Proceedings of the International Conference “Health and wellbeing in modern society” (ICHW 2020) PB - Atlantis Press SP - 95 EP - 99 SN - 2468-5739 UR - https://doi.org/10.2991/ahsr.k.201001.020 DO - 10.2991/ahsr.k.201001.020 ID - Bobkov2020 ER -