Study of Wear on the Metal on Polycarbonate Urethane (PCU) Contact in Total Hip Arthroplasty Using Finite Element Method
- DOI
- 10.2991/aer.k.210810.022How to use a DOI?
- Keywords
- PCU, Contact Pressure, Von Mises Stress, FEM
- Abstract
Latterly, the polycarbonare urethane (PCU) has suggested a viable substitude to conventional bearings. The aim of this study is to determine the von Mises stress and contact pressure as a function of different of acetabular cup thickness. The analysis of this study was conducted by the finite element method. Six variation of acetabular liner thickness (5mm, 10mm, 15mm, 20mm, 30mm, and 40mm) were used in this simulation. The contact pressure was determined to predict the wear performance of PCU acetabular liner. The result shown that the thicker of the acetabular liner, the smaller the contact pressure and the smaller the contact radius. Thus predictable that the thinner the PCU acetabular liner, the higher the wear rate.
- Copyright
- © 2021, the Authors. Published by Atlantis Press.
- Open Access
- This is an open access article distributed under the CC BY-NC license (http://creativecommons.org/licenses/by-nc/4.0/).
Cite this article
TY - CONF AU - Wahyu Dwi Lestari AU - Luluk Edahwati AU - Ndaru Adyono AU - Rifky Ismail AU - J Jamari AU - Athanasius Priharyoto Bayuseno PY - 2021 DA - 2021/08/11 TI - Study of Wear on the Metal on Polycarbonate Urethane (PCU) Contact in Total Hip Arthroplasty Using Finite Element Method BT - Proceedings of the 2nd Borobudur International Symposium on Science and Technology (BIS-STE 2020) PB - Atlantis Press SP - 137 EP - 143 SN - 2352-5401 UR - https://doi.org/10.2991/aer.k.210810.022 DO - 10.2991/aer.k.210810.022 ID - Lestari2021 ER -