Artery Research

Volume 7, Issue 3-4, September 2013, Pages 168 - 168

4.5 FREQUENCY RESPONSE OF BLOOD PRESSURE CUFFS BASED ON STEP RESPONSE AND FORCED SINUSOIDAL HARMONIC EXCITATION

Authors
R. Lurf1, R. Semerad1, M. Meindl1, C. Mayer1, B. Hametner1, T. Weber2, S. Wassertheurer1
1AIT Austrian Institute of Technology, Vienna, Austria
2Klinikum Wels-Grieskirchen, Wels, Austria
Available Online 11 November 2013.
DOI
10.1016/j.artres.2013.10.023How to use a DOI?
Abstract

Background. Pulse wave analysis (PWA) using cuff based methods emerged within the last years. Compared to traditional sensors used in PWA like catheters and piezo gauges very little is known on the frequency response of cuff based systems according to standard measurement and systems theory. Therefore the aim of this work is the investigation of the capability of blood pressure cuffs to register the dynamical behavior of the arterial pulse appropriately.

Methods. To evaluate the dynamic behavior of cuff based sensor chains we performed both tests on step response (similar to “pop test”) as well as forced sinusoidal harmonic excitation by the means of a fully automated and standardized custom testing bench. The stepwise variation of cuff pressures and/or bladder volumes and cuff sizes was intended to account for various anatomical situations faced in clinical practice.

Results. The resonant frequencies of the evaluated systems are in the range of 110 Hz (Fig. 1a). Similar results have been obtained for both step response method and harmonic excitation. This behavior did not vary over a pressure range from 80 to 200 mmHg. Nevertheless we observed that pressure-volume relation had significant influence on the cuffs frequency response. In particular too tightly fitted cuffs lead to chaotic results (Fig. 1b). Too loosely fitted cuffs causes a loss in resolution.

Conclusions. Based on our actual data blood pressure cuffs provide acceptable capabilities to cope with the dynamics of the arterial pulse for PWA use. This is only valid for properly fitted and inflated cuffs.

Open Access
This is an open access article distributed under the CC BY-NC license.

Journal
Artery Research
Volume-Issue
7 - 3-4
Pages
168 - 168
Publication Date
2013/11/11
ISSN (Online)
1876-4401
ISSN (Print)
1872-9312
DOI
10.1016/j.artres.2013.10.023How to use a DOI?
Open Access
This is an open access article distributed under the CC BY-NC license.

Cite this article

TY  - JOUR
AU  - R. Lurf
AU  - R. Semerad
AU  - M. Meindl
AU  - C. Mayer
AU  - B. Hametner
AU  - T. Weber
AU  - S. Wassertheurer
PY  - 2013
DA  - 2013/11/11
TI  - 4.5 FREQUENCY RESPONSE OF BLOOD PRESSURE CUFFS BASED ON STEP RESPONSE AND FORCED SINUSOIDAL HARMONIC EXCITATION
JO  - Artery Research
SP  - 168
EP  - 168
VL  - 7
IS  - 3-4
SN  - 1876-4401
UR  - https://doi.org/10.1016/j.artres.2013.10.023
DO  - 10.1016/j.artres.2013.10.023
ID  - Lurf2013
ER  -