Abstract

Introduction: The assessment of middle cerebral artery (MCA) haemodynamics is essential for the diagnosis and monitoring of cerebrovascular disease. However, conventional transcranial Doppler (TCD) may not capture the correct flow velocities because of sub-optimal angles of insonation. Conversely, 2-D ultrasound (2D-US) allows for the visualisation and angle-correction of MCA haemodynamics. Therefore, this study aimed to determine potential differences in MCA haemodynamics obtained with TCD and 2D-US.

Methods: MCA haemodynamics were obtained in a blinded, randomised order with TCD and 2D-US (non-angle-corrected = 2D-US-NON and angle-corrected = 2D-US-ANGLE) from the temporal left posterior window in twenty-seven healthy participants. Recordings were analysed for peak-systolic velocity (PSV), end-diastolic velocity (ED), pulsatility index (PI) and resistance index (RI). Statistical agreements between TCD and 2D-US were determined using linear regression, independent samples t-test and Bland-Altman analysis.

Results: MCA haemodynamics obtained with TCD explained less than 50% of the values obtained with 2D-US-NON & 2D-US-ANGLE, respectively (PSV r2 = 0.34 & 0.37; ED: r2 = 0.37 & 0.44; PI: r2 = 0.20 & 0.22; RI: r2 = 0.30 & 0.32). Compared with 2D-US-NON, TCD produced similar PSV (p = 0.65) but significantly higher ED (p < 0.0001), lower PI (p < 0.0001) and lower RI (p < 0.0001). 2D-US angle-correction resulted in a significantly higher PSV compared with TCD (p = 0.005) while all other differences remained. Bland-Altman analysis revealed a bias between the two methods ranging from 11–40%, with large individual variability.

Conclusion: TCD and 2D ultrasound produce significantly different values for MCA haemodynamics, even when 2D-US is non-angle-corrected. This may have important implications when using indices of MCA haemodynamics in the evaluation of cerebrovascular disease.

Copyright
© 2019 Association for Research into Arterial Structure and Physiology. Publishing services by Atlantis Press International B.V.
Open Access
This is an open access article distributed under the CC BY-NC 4.0 license (http://creativecommons.org/licenses/by-nc/4.0/).

Journal
Artery Research
Volume-Issue
25 - Supplement 1
Pages
S100 - S100
Publication Date
2020/02/17
ISSN (Online)
1876-4401
ISSN (Print)
1872-9312
DOI
10.2991/artres.k.191224.090How to use a DOI?
Copyright
© 2019 Association for Research into Arterial Structure and Physiology. Publishing services by Atlantis Press International B.V.
Open Access
This is an open access article distributed under the CC BY-NC 4.0 license (http://creativecommons.org/licenses/by-nc/4.0/).

Cite this article

TY  - JOUR
AU  - Christopher Brown
AU  - Mahfoudha Al Shezawi
AU  - Laura Watkeys
AU  - Maggie Munnery
AU  - Christopher Pugh
AU  - Eric Stöhr
AU  - Barry McDonnell
PY  - 2020
DA  - 2020/02/17
TI  - P59 Marked Differences in Cerebral Haemodynamics Obtained with Transcranial Doppler vs. 2-D Angle-corrected Ultrasound
JO  - Artery Research
SP  - S100
EP  - S100
VL  - 25
IS  - Supplement 1
SN  - 1876-4401
UR  - https://doi.org/10.2991/artres.k.191224.090
DO  - 10.2991/artres.k.191224.090
ID  - Brown2020
ER  -