Artery Research

Volume 25, Issue Supplement 1, December 2019, Pages S153 - S153

P113 The Forgotten Black Box of Blood Pressure: Error in Oscillometric Mean Arterial Pressure is Associated with Cuff Measurement Inaccuracy

Authors
Dean Picone1, Martin Schultz1, Matthew Armstrong1, Nathan Dwyer1, 2, Philip Roberts-Thomson1, 2, Thomas Weber3, James Sharman2
1Menzies Institute for Medical Research, College of Health and Medicine, University of Tasmania, Australia
2Royal Hobart Hospital, Hobart, Australia
3Cardiology Department, Klinikum Wels-Grieskirchen, Wels, Austria
Available Online 17 February 2020.
DOI
10.2991/artres.k.191224.139How to use a DOI?
Abstract

Background: Accurate measurement of cuff systolic BP (SBP) and diastolic BP (DBP) is contingent on the assumption of oscillometric MAP being accurate. If MAP is inaccurate this is likely to impact the accuracy of cuff SBP and DBP, but this has never been determined and was the aim of the study.

Methods: In five independent studies, MAP was measured by five unique cuff oscillometric BP devices at the same time as invasive aortic MAP (area under the waveform) by fluid-filled or solid-state catheters among 262 patients (61 ± 11 years, 65% male) during coronary angiography. Cuff SBP and DBP were estimated by device-specific algorithms. Measurement errors were calculated as cuff – invasive aortic BP.

Results: For each BP device, MAP error was −2.6 ± 6.1, 0.76 ± 7.2, 1.4 ± 5.5, 1.5 ± 6.7 and 7.1 ± 9.2 mmHg. Corresponding SBP errors were 4.4 ± 8.7, −0.8 ± 11.2, −8.8 ± 10.4, −0.9 ± 10.5 and 1.9 ± 10.8 mmHg, whereas DBP errors were 8.8 ± 5.1, 2.0 ± 7.4, 6.7 ± 7.3, 10.3 ± 9.0 and 13.1 ± 6.4 mmHg. There was a direct relationship between errors in oscillometric MAP and SBP error in four of the five devices (B range = 0.42 to 1.04). However, MAP error was consistently related to DBP error in all devices (B range = 0.48 to 0.97). Across all devices, absolute error in MAP ≥3 mmHg corresponded to absolute error in SBP and DBP ≥5 mmHg in 56–77% and 62–88% of patients.

Conclusion: Errors in oscillometric MAP are directly related to cuff SBP and DBP inaccuracy, but the magnitude of error is device-specific. Further work is required to understand algorithms used in oscillometric devices to improve cuff BP accuracy.

Copyright
© 2019 Association for Research into Arterial Structure and Physiology. Publishing services by Atlantis Press International B.V.
Open Access
This is an open access article distributed under the CC BY-NC 4.0 license (http://creativecommons.org/licenses/by-nc/4.0/).

Journal
Artery Research
Volume-Issue
25 - Supplement 1
Pages
S153 - S153
Publication Date
2020/02/17
ISSN (Online)
1876-4401
ISSN (Print)
1872-9312
DOI
10.2991/artres.k.191224.139How to use a DOI?
Copyright
© 2019 Association for Research into Arterial Structure and Physiology. Publishing services by Atlantis Press International B.V.
Open Access
This is an open access article distributed under the CC BY-NC 4.0 license (http://creativecommons.org/licenses/by-nc/4.0/).

Cite this article

TY  - JOUR
AU  - Dean Picone
AU  - Martin Schultz
AU  - Matthew Armstrong
AU  - Nathan Dwyer
AU  - Philip Roberts-Thomson
AU  - Thomas Weber
AU  - James Sharman
PY  - 2020
DA  - 2020/02/17
TI  - P113 The Forgotten Black Box of Blood Pressure: Error in Oscillometric Mean Arterial Pressure is Associated with Cuff Measurement Inaccuracy
JO  - Artery Research
SP  - S153
EP  - S153
VL  - 25
IS  - Supplement 1
SN  - 1876-4401
UR  - https://doi.org/10.2991/artres.k.191224.139
DO  - 10.2991/artres.k.191224.139
ID  - Picone2020
ER  -