P76 PERIPHERAL BLOOD PRESSURE AND STIFFNESS INDEX ESTIMATION USING THE PPG SIGNAL MEASURED WITH THE pOpmètre DEVICE: CALIBRATION WITH THE CUFF BLOOD PRESSURE
- DOI
- 10.1016/j.artres.2017.10.092How to use a DOI?
- Abstract
Arterial stiffness may influence the contour of the peripheral pulse, suggesting that contour analysis of the digital volume pulse (DVP) might be used to estimate peripheral blood pressure and a stiffness index (SI).
The objective is to establish a transfer function that estimates the peripheral blood pressure using the PPG pulse, calibrated with a brachial pressure cuff, and then to deduce a stiffness index using the established parameters.
We positioned the photodiode sensor on the finger. Brachial blood pressure measurement was performed with a cuff adapted to the arm circumference and an oscillometric device (Omron M10 -IT). Pulse wave velocity (PWV) measurements were performed with the pOpmètre® system. DVP waveforms were recorded over a 10 s period and ensemble-averaged to obtain a single waveform from which DT (DVP) was determined as the time between the first systolic peak and the early diastolic peak/inflection point in the waveform. The SI is the ratio of the subject’s height to the DT (DVP).
69 subjects were included: 24 healthy subjects and 45 patients with essential hypertension. The correlation between the estimated peripheral diastolic pressure and the brachial one was good and significant (r2 = 0.51; p < 0.001). A better correlation was found in terms of peripheral systolic pressure (r2 = 0.56; p < 0.001). The correlation between the SI and the ft-PWV was significant (r = 0.5; p < 0.001) classifying the estimation as good agreement.
The estimation of the peripheral blood pressure and a stiffness index with the PPG signal qualifies as good agreement with the reference technique.
- Open Access
- This is an open access article distributed under the CC BY-NC license.
Cite this article
TY - JOUR AU - Hasan Obeid AU - Hakim Khettab AU - Magid Hallab AU - Stephane Laurent AU - Pierre Boutouyrie PY - 2017 DA - 2017/12/06 TI - P76 PERIPHERAL BLOOD PRESSURE AND STIFFNESS INDEX ESTIMATION USING THE PPG SIGNAL MEASURED WITH THE pOpmètre DEVICE: CALIBRATION WITH THE CUFF BLOOD PRESSURE JO - Artery Research SP - 74 EP - 74 VL - 20 IS - C SN - 1876-4401 UR - https://doi.org/10.1016/j.artres.2017.10.092 DO - 10.1016/j.artres.2017.10.092 ID - Obeid2017 ER -