2.5 COMPARISON OF NON-INVASIVE AND INVASIVE MEASUREMENTS OF CENTRAL BLOOD PRESSURE IN PATIENTS WITH CHRONIC KIDNEY DISEASE

Authors
R. Carlsen, C. Peters, D. Khatir, E. Laugesen, S. Winther, N. Buus
Aarhus University Hospital, Aarhus, Denmark
Available Online 4 November 2014.
DOI
10.1016/j.artres.2014.09.061How to use a DOI?
Open Access
This is an open access article distributed under the CC BY-NC license.

Background: The blood pressure (BP) in the ascending aorta can be estimated non-invasively by pulse wave analysis using commercially available devices such as the SphygmoCor. However, this device has not been validated for use in patients with chronic kidney disease (CKD).

Objectives: Comparison of non-invasively obtained estimates of central BP with invasively measured central BP in CKD patients.

Methods: Patients with stable CKD stage 3–5 corresponding to an estimated glomerular filtration rate (eGFR) under 60 ml/min undergoing elective coronary angiography were included. Central BP was estimated by the SphygmoCor using four different calibrations: Invasive aortic diastolic and mean BP (invasive), systolic and diastolic brachial BP (sys-dia), brachial diastolic and mean arterial BP (form factor 0.33) (ff-0.33), and brachial diastolic and mean arterial BP (form factor 0.4) (ff-0.4). Brachial artery BP was measured simultaneously with a validated oscillometric BP device.

Results: Forty-nine patients, 65% males, age 61±13 (mean±SD) with a median eGFR of 18 ml/min (range 5–59 ml/min) were enrolled. Invasive BP was 150.8±22/ 75.4±11 mmHg while brachial BP was 147.3±19 /86.0±10 mmHg (mean differences: −3.6±9.3/10.6±7.9 mmHg (P<0.012).

Mean differences (invasive minus estimated central BP) with the four calibrations used were: −5.9±6.1/1.3±1.0 mmHg, P<0.001 (invasive); −16.0±9.3/11.5±8.0 mmHg, P<0.001 (sys-dia); −17.5±10.5/11.5±7.9 mmHg, P<0.001 (ff-0.33); −7.5±10.2 /11.6±7.9 mmHg, P<0.001 (ff-0.4).

Conclusion: In CKD patients, we found a systematic difference between estimated and invasively measured central BP. Surprisingly, brachial systolic BP was very close to invasively measured central systolic BP and was even more accurate than estimates based on calibration with invasively obtained BP.

Journal
Artery Research
Volume-Issue
8 - 4
Pages
124 - 124
Publication Date
2014/11/04
ISSN (Online)
1876-4401
ISSN (Print)
1872-9312
DOI
10.1016/j.artres.2014.09.061How to use a DOI?
Open Access
This is an open access article distributed under the CC BY-NC license.

Cite this article

TY  - JOUR
AU  - R. Carlsen
AU  - C. Peters
AU  - D. Khatir
AU  - E. Laugesen
AU  - S. Winther
AU  - N. Buus
PY  - 2014
DA  - 2014/11/04
TI  - 2.5 COMPARISON OF NON-INVASIVE AND INVASIVE MEASUREMENTS OF CENTRAL BLOOD PRESSURE IN PATIENTS WITH CHRONIC KIDNEY DISEASE
JO  - Artery Research
SP  - 124
EP  - 124
VL  - 8
IS  - 4
SN  - 1876-4401
UR  - https://doi.org/10.1016/j.artres.2014.09.061
DO  - 10.1016/j.artres.2014.09.061
ID  - Carlsen2014
ER  -