Research on Environmental Test On-line Detection Device Based on Flow Metrological Instrument
- DOI
- 10.2991/msmee-17.2017.253How to use a DOI?
- Keywords
- flow metrology meter; impact of ambient change; online dynamic measurement
- Abstract
The metrological performance of flow meter should be tested under the condition of ambient temperature and humidity changes according to appendix A type evaluation outline of JJG 1033-2007 "Electromagnetic Flowmeter" verification procedures. In order to meet the requirements of the type of evaluation outline, an "environmental test on-line detection device based on flow metrological instrument" is developed. The pipe and the meter can be directly connected to this device. The metrological performance of the meter can be assessed and the environment's temperature and humidity can be controlled at the same time. Experts have identified that the device can fully meet the requirement that metrological performance of flow meter can be tested under the condition of ambient temperature and humidity changes.
- Copyright
- © 2017, the Authors. Published by Atlantis Press.
- Open Access
- This is an open access article distributed under the CC BY-NC license (http://creativecommons.org/licenses/by-nc/4.0/).
Cite this article
TY - CONF AU - Fanliang Meng AU - Senlin Gao AU - Quan Sun AU - Xiaoping Zhang AU - Xigang Wang PY - 2017/05 DA - 2017/05 TI - Research on Environmental Test On-line Detection Device Based on Flow Metrological Instrument BT - Proceedings of the 2017 2nd International Conference on Materials Science, Machinery and Energy Engineering (MSMEE 2017) PB - Atlantis Press SP - 1371 EP - 1375 SN - 2352-5401 UR - https://doi.org/10.2991/msmee-17.2017.253 DO - 10.2991/msmee-17.2017.253 ID - Meng2017/05 ER -