Research on the Parameters of HIFU Based on Near-field Cross-spectrum Method
- DOI
- 10.2991/mmsa-18.2018.74How to use a DOI?
- Keywords
- HIFU technology; near-field cross-spectrum method; sound intensity; sound power
- Abstract
This paper aims at the measurement problem of High Intensity Focused Ultrasound (HIFU). In order to measure the sound parameters of HIFU, we study the near-field cross-spectrum method. That is to say, the complex sound pressure of two adjacent planes can be measured in the near field, besides, the sound power and the sound intensity can be obtained through the cross-spectrum relationship between two groups of sound pressure. For verifying the theory, we used a self-built underwater acoustic field measurement system to compare the far-field method with the near-field cross-spectrum method by measuring the sound power of the plane piston transducer, and the result shows that the near-field cross-spectrum method is reliable in the effective range of acoustic error.
- Copyright
- © 2018, the Authors. Published by Atlantis Press.
- Open Access
- This is an open access article distributed under the CC BY-NC license (http://creativecommons.org/licenses/by-nc/4.0/).
Cite this article
TY - CONF AU - Zhuohan Tang AU - Huifeng Zheng AU - Jiayu Gao AU - Linjing Wu AU - Yanchao Wang PY - 2018/03 DA - 2018/03 TI - Research on the Parameters of HIFU Based on Near-field Cross-spectrum Method BT - Proceedings of the 2018 International Conference on Mathematics, Modelling, Simulation and Algorithms (MMSA 2018) PB - Atlantis Press SP - 335 EP - 339 SN - 1951-6851 UR - https://doi.org/10.2991/mmsa-18.2018.74 DO - 10.2991/mmsa-18.2018.74 ID - Tang2018/03 ER -