Research on Parameters’ Determination and Imaging of CT System
- DOI
- 10.2991/mmsa-18.2018.52How to use a DOI?
- Keywords
- filtered back projection algorithm; radon inverse transformation; tomography; CT system parameters; permutation function
- Abstract
In this paper, to determine parameters of the CT system and reconstruct images, the geometric model of the template is used to calculate its parameters, and the image is reconstructed based on the Filtered back projection algorithm, which contains the RADON inverse transform and the usage of calibrated CT system parameters. To remove the noise of the image background, we optimize the image of CT with discrete properties of pixels. What is more, this paper presents a new template that is more appropriate to determine CT system parameters. And it is more conducive to calculating the unit spacing and the receiving angle of detectors, and therefore this new system improves the calibration accuracy of CT system parameters.
- Copyright
- © 2018, the Authors. Published by Atlantis Press.
- Open Access
- This is an open access article distributed under the CC BY-NC license (http://creativecommons.org/licenses/by-nc/4.0/).
Cite this article
TY - CONF AU - Nan Qiao AU - Jiu Xiong AU - Qi Huang PY - 2018/03 DA - 2018/03 TI - Research on Parameters’ Determination and Imaging of CT System BT - Proceedings of the 2018 International Conference on Mathematics, Modelling, Simulation and Algorithms (MMSA 2018) PB - Atlantis Press SP - 229 EP - 232 SN - 1951-6851 UR - https://doi.org/10.2991/mmsa-18.2018.52 DO - 10.2991/mmsa-18.2018.52 ID - Qiao2018/03 ER -