Allocation of Fetal Heart Tones During Non-stop Monitoring of Pregnant Women
- DOI
- 10.2991/itsmssm-16.2016.65How to use a DOI?
- Keywords
- cardiotocography, phonocardiography, allocation of heart tones separation, daily monitoring
- Abstract
The article is to study the system of non-stop daily monitoring of the state of the maternal and the fetal circulatory systems using acoustic sensors. After preliminary preparation of the phonocardiogram heart tones of the fetus and the pregnant woman were extracted using Shannon energy. The obtained data was separated, followed by calculation of the heart beats. As a result, software algorithms for calculating heart beats at a particular time (30 seconds to 6 hours) have been developed and implemented. The measurement error ranges from 1 to 6 beats per minute. The accuracy of the software and algorithmic support was estimated by comparing the results obtained during parallel cardiotocography, taken as the reference method.
- Copyright
- © 2016, the Authors. Published by Atlantis Press.
- Open Access
- This is an open access article distributed under the CC BY-NC license (http://creativecommons.org/licenses/by-nc/4.0/).
Cite this article
TY - CONF AU - Ekaterina Kiseleva AU - Dmitriy Zhdanov AU - Ivan Zemlyakov AU - Elena Vaganova AU - Sergey Yuryev PY - 2016/05 DA - 2016/05 TI - Allocation of Fetal Heart Tones During Non-stop Monitoring of Pregnant Women BT - Proceedings of the 2016 Conference on Information Technologies in Science, Management, Social Sphere and Medicine PB - Atlantis Press SP - 321 EP - 324 SN - 2352-538X UR - https://doi.org/10.2991/itsmssm-16.2016.65 DO - 10.2991/itsmssm-16.2016.65 ID - Kiseleva2016/05 ER -