The Calibration of Wide Range Meter Based on ML610Q495
Authors
Nan Wang
Corresponding Author
Nan Wang
Available Online January 2015.
- DOI
- 10.2991/isci-15.2015.101How to use a DOI?
- Keywords
- Wide Range Meter; Error Calibration; High Accuracy RTC; RTC Calibration
- Abstract
According to the problem that introduces the design of smart meters,which is based on ML610Q495 microcontroller and IDT90E23 measurement chip. This meter has the characteristics of wide range measuring and high accuracy RTC. In the range of 0.02A~100A current, the meter’s measuring accuracy is below ±0.2%, and in the range of -45 C~+85 C temperature the meter’s RTC accuracy is below ±0.1s/d. This design can meet all the requirements of the State Grid, China Southern Power Grid, and Neimenggu Power Grid, and the meters have passed the test.
- Copyright
- © 2015, the Authors. Published by Atlantis Press.
- Open Access
- This is an open access article distributed under the CC BY-NC license (http://creativecommons.org/licenses/by-nc/4.0/).
Cite this article
TY - CONF AU - Nan Wang PY - 2015/01 DA - 2015/01 TI - The Calibration of Wide Range Meter Based on ML610Q495 BT - Proceedings of the 2015 International Symposium on Computers & Informatics PB - Atlantis Press SP - 756 EP - 763 SN - 2352-538X UR - https://doi.org/10.2991/isci-15.2015.101 DO - 10.2991/isci-15.2015.101 ID - Wang2015/01 ER -