Design of a weak fluorescence detection circuit in molecular diagnosis
- DOI
- 10.2991/iiicec-15.2015.413How to use a DOI?
- Keywords
- Molecular diagnosis; weak fluorescence detection; photomultiplier tubes
- Abstract
Fluorescence is generally adopted in molecular diagnosis. In order to detect weak fluorescence signal, a weak fluorescence detection circuit has been design based on a photomultiplier tube. The circuit mainly consists of three parts which are a photoelectric converter circuit, a pre-amplifier circuit and a filter circuit. This circuit is able to receive weak fluorescent signals and convert them into electrical signals effectively. As a result, important information such as the concentration of corresponding reagnet can be obtained by analyzing the intensity of fluorescence received. It is experimentally demonstrated that the designed circuit can be used in weak fluorescence detecting efficiently. Its detection limitation is as low as the fluorescent signal emitted from quantum dots of 10-10mol/ml, making it suitable for low fluorescence detection in molecular diagnosis.
- Copyright
- © 2015, the Authors. Published by Atlantis Press.
- Open Access
- This is an open access article distributed under the CC BY-NC license (http://creativecommons.org/licenses/by-nc/4.0/).
Cite this article
TY - CONF AU - Huajuan Bai AU - Jihong Feng AU - Hong Zhang PY - 2015/03 DA - 2015/03 TI - Design of a weak fluorescence detection circuit in molecular diagnosis BT - Proceedings of the 2015 International Industrial Informatics and Computer Engineering Conference PB - Atlantis Press SP - 1899 EP - 1902 SN - 2352-538X UR - https://doi.org/10.2991/iiicec-15.2015.413 DO - 10.2991/iiicec-15.2015.413 ID - Bai2015/03 ER -