The circuit design for extraction of electro encephalogram signals
- DOI
- 10.2991/ifeesd-16.2016.80How to use a DOI?
- Keywords
- EEG signals; amplification; Wave filtering; Circuit Design
- Abstract
The design of front-end signals acquisition circuit is made for extraction of EEG signals, thus making it possible to amplify, filter, and remove the interference of source signal collected from the brain electrode. After the signal go in turn through such links as pre-amplification, high-pass filtration, 50Hz wave trapping, low-pass filtration, and post amplification, the EEG signals average amplitude of 50 V can reach to between 1V ~4V, thereby meeting the range of requirements the conventional acquisition card needs for the input of analog signal. The simulation results show that the signal, as collected through the circuit, are low in distortion, and thus can be widely used in the devices for processing EEG signals.
- Copyright
- © 2016, the Authors. Published by Atlantis Press.
- Open Access
- This is an open access article distributed under the CC BY-NC license (http://creativecommons.org/licenses/by-nc/4.0/).
Cite this article
TY - CONF AU - Wenqiang Liu AU - Lijun Qiu AU - Qifu Fan AU - Huan Liu PY - 2016/05 DA - 2016/05 TI - The circuit design for extraction of electro encephalogram signals BT - Proceedings of the 2016 International Forum on Energy, Environment and Sustainable Development PB - Atlantis Press SP - 453 EP - 457 SN - 2352-5401 UR - https://doi.org/10.2991/ifeesd-16.2016.80 DO - 10.2991/ifeesd-16.2016.80 ID - Liu2016/05 ER -