Detection and analysis of 18 PAHs in water by micro-LLE combined with concurrent solvent recondensation-large volume injection-GC-MS
- DOI
- 10.2991/icsmim-15.2016.138How to use a DOI?
- Keywords
- Detection, 18 PAHs, micro-LLE, GC-MS, recondensation-large volume injection
- Abstract
A rapid and efficient method termed micro liquid-liquid extraction combined with concurrent solvent recondensation-large volume injection-GC-MS has been developed for the extraction and determination of PAHs in water. Some condition parameters were optimized to obtain high detection sensitivity. Calibration curves of 18 PAHs were linear in the range of 0.1-10 ng•L 1, with the correlation coefficients (r2) between 0.9993 and 1.0. The method attained good precisions from 0.3% to 4.5% for PAHs standard solutions. The method recoveries ranged between 64.5-119.5% for spiked with 5.0 ng•L 1 and 76.2-119.5% for spiked with 25.0 ng•L 1 and 50.0 ng•L 1 for real water samples. Limits of detection (LODs, S/N=3) of the method were determined from 0.010-0.068 ng•L 1. The optimized method was successfully applied to the determination of 18 PAHs in real water samples.
- Copyright
- © 2016, the Authors. Published by Atlantis Press.
- Open Access
- This is an open access article distributed under the CC BY-NC license (http://creativecommons.org/licenses/by-nc/4.0/).
Cite this article
TY - CONF AU - Xiaolin Zhou AU - Bo Xu AU - Liya Cao AU - Xinying Zhang PY - 2016/01 DA - 2016/01 TI - Detection and analysis of 18 PAHs in water by micro-LLE combined with concurrent solvent recondensation-large volume injection-GC-MS BT - Proceedings of the 2015 4th International Conference on Sensors, Measurement and Intelligent Materials PB - Atlantis Press SP - 736 EP - 739 SN - 2352-538X UR - https://doi.org/10.2991/icsmim-15.2016.138 DO - 10.2991/icsmim-15.2016.138 ID - Zhou2016/01 ER -