- Abstract
This Medical needle plays an important role in modern clinical medicine. The penetration and pull process of medical injection needle was simulated by FEM software Abaqus, and the equivalent stress and strain were analyzed. Self-developed mechanical loading equipment was adopted to test the penetration force with different needle tip angles, needle diameters and penetration speeds. The results showed that: the puncture of medical needle into skin was mode I crack. The maximum stress and strain was focused on skin around the tip of needle. The penetration force decreased with needle diameter and penetration speed. The angle of needle tip form greatly affected the penetration force. When the angle of needle tip was 90°, the penetration force reached a maximum value of 708.3 mN, which was 2.21 times that of the needle tip angle 30°. The penetration force of isosceles triangle tip was minimum value.
- Copyright
- © 2016, the Authors. Published by Atlantis Press.
- Open Access
- This is an open access article distributed under the CC BY-NC license (http://creativecommons.org/licenses/by-nc/4.0/).
Cite this article
TY - CONF AU - Lelun Jiang AU - Yong Huang AU - Chengfeng Pan AU - Jintian Ling AU - Jiyu Li AU - Zhipeng Chen AU - Jilin Fu AU - Yubiao Wen PY - 2016/04 DA - 2016/04 TI - Research on Insertion Process of Medical Needle BT - Proceedings of the 2016 3rd International Conference on Materials Engineering, Manufacturing Technology and Control PB - Atlantis Press SP - 779 EP - 785 SN - 2352-5401 UR - https://doi.org/10.2991/icmemtc-16.2016.154 DO - 10.2991/icmemtc-16.2016.154 ID - Jiang2016/04 ER -