Wireless Charging Device for Artificial Cardiac Pacemaker
Authors
Quan Xiong
Corresponding Author
Quan Xiong
Available Online October 2015.
- DOI
- 10.2991/icitmi-15.2015.127How to use a DOI?
- Keywords
- wireless power transfer; magnetic-coupled; cardiac pacemaker; battery charging
- Abstract
To overcome the limitation of energy supply for artificial cardiac pacemaker, this paper investigated a wireless charging device based on the magnetic coupling resonant wireless transfer (MCR-WPT) technique. We designed the bilayer coupled coils for energy transmission, which had twice power factor than monolayer coils. The frequency is 200 kHz when the MCR-WPT system was in resonance. Furthermore, we put the titanium alloy case behind the receiver coil to simulate the environment of artificial cardiac pacemaker. At last, we have done the charging experiment.
- Copyright
- © 2015, the Authors. Published by Atlantis Press.
- Open Access
- This is an open access article distributed under the CC BY-NC license (http://creativecommons.org/licenses/by-nc/4.0/).
Cite this article
TY - CONF AU - Quan Xiong PY - 2015/10 DA - 2015/10 TI - Wireless Charging Device for Artificial Cardiac Pacemaker BT - Proceedings of the 4th International Conference on Information Technology and Management Innovation PB - Atlantis Press SP - 765 EP - 768 SN - 2352-538X UR - https://doi.org/10.2991/icitmi-15.2015.127 DO - 10.2991/icitmi-15.2015.127 ID - Xiong2015/10 ER -