Study on Photoelectric Sensor to Detect the Concentration of SO42-
- DOI
- 10.2991/icadme-15.2015.81How to use a DOI?
- Keywords
- electrolyte, concentration detection, photoelectric sensor, SO42-
- Abstract
We’ve designed a photoelectric sensor to detect the concentration of SO42- solution with a vertical capsule which was immersed into sample solution derectly. The system contains a pair of light-emitting diode (LED) and photosensitive diode (PIN) packaged with cover of semi-sphere shape. Theoretical analysis and experimental results have shown that the relative transmission rate of the system was almost proportional to the concentration of solution; which meant that the more concentrated the solution was, the less the relative transmission rate would be. In our experiments, the concentration of SO42- solution can be detected within the range of 0-3.441mol/L by using a LED with a central wavelength of 940nm. We have also confirmed that the concentration of Zn2+ ( 3.441 mol/L), Cu2+ ( 0.5627 mol/L) and H+ ( 2.015 mol/L) had tiny effects on the detection results of SO42- concentration.
- Copyright
- © 2015, the Authors. Published by Atlantis Press.
- Open Access
- This is an open access article distributed under the CC BY-NC license (http://creativecommons.org/licenses/by-nc/4.0/).
Cite this article
TY - CONF AU - Ni Feng AU - Ni Ruodan PY - 2015/10 DA - 2015/10 TI - Study on Photoelectric Sensor to Detect the Concentration of SO42- BT - Proceedings of the 5th International Conference on Advanced Design and Manufacturing Engineering PB - Atlantis Press SP - 403 EP - 409 SN - 2352-5401 UR - https://doi.org/10.2991/icadme-15.2015.81 DO - 10.2991/icadme-15.2015.81 ID - Feng2015/10 ER -