Accelerometer Static Calibration based on the PSO algorithm
- DOI
- 10.2991/emeit.2012.119How to use a DOI?
- Keywords
- accelerometer, static calibration, the PSO algorithm
- Abstract
This paper introduces MEMS accelerometer static calibration method based on the PSO algorithm. According to the mathematical model of the MEMS accelerometer, the PSO algorithm is applied to estimate the mathematical model, including scale factors, bias, and installation errors. The method does not require a three-axis rotary platform, and the calculation is simple, easy to implement, high accuracy and provides a new choice for outdoor testing environment. Analyzing the experimental data, the random errors in the sampling process have been eliminated effectively, and the calibration effect is better than the commonly static correction method.
- Copyright
- © 2012, the Authors. Published by Atlantis Press.
- Open Access
- This is an open access article distributed under the CC BY-NC license (http://creativecommons.org/licenses/by-nc/4.0/).
Cite this article
TY - CONF AU - Liguo Zhang AU - Wenchao Li AU - Huilian Liu PY - 2012/09 DA - 2012/09 TI - Accelerometer Static Calibration based on the PSO algorithm BT - Proceedings of the 2nd International Conference on Electronic & Mechanical Engineering and Information Technology (EMEIT 2012) PB - Atlantis Press SP - 570 EP - 573 SN - 1951-6851 UR - https://doi.org/10.2991/emeit.2012.119 DO - 10.2991/emeit.2012.119 ID - Zhang2012/09 ER -