P6.15 HAEMODYNAMIC-MECHANICAL INTERACTION IN TYPE-B AORTIC DISSECTION: EXPERIMENTAL AND COMPUTER MODEL STUDY
- DOI
- 10.1016/j.artres.2013.10.196How to use a DOI?
- Abstract
Background: Type-B aortic dissections (TBADs) show a high morbidity and mortality in the long-term follow up. The role of and interaction between haemodynamics (pressure and flows), mechanics (wall compliance) and morphology (communication between false lumen (FL) and true lumen (TL)) is poorly understood, as well as the potential role of these variables in the outcome of TBADs.
Materials and methods: We developed both a hydraulic bench (experimental) and an electrical analogue model of a TBAD (Fig. 1). The model was calibrated and validated for eight experimental cases consisting of various permutations of proximal and distal tear sizes (no tear, 4 and 10 mm).
Results: The computer model was able to reproduce measured intraluminal pressures and velocities across tears with good agreement for all combinations (Fig. 2). Tear size is the major determinant of interaction between FL and TL. Tear sizes of 4 mm are associated with a reduced pressure in the FL, and strong oscillatory flow exchange between TL and FL. A tear size of 10 mm increases FL pressure to the same level or beyond TL pressures, and reduces the oscillatory nature of the flow.
Conclusions: Both the experimental and computer model reveal a complex dynamic interaction between TL and FL, modulated by the size and combination of the tears.
- Open Access
- This is an open access article distributed under the CC BY-NC license.
Cite this article
TY - JOUR AU - P.A. Rudenick AU - B.H. Bijnens AU - P. Segers AU - D. García-Dorado AU - A. Evangelista PY - 2013 DA - 2013/11/11 TI - P6.15 HAEMODYNAMIC-MECHANICAL INTERACTION IN TYPE-B AORTIC DISSECTION: EXPERIMENTAL AND COMPUTER MODEL STUDY JO - Artery Research SP - 158 EP - 158 VL - 7 IS - 3-4 SN - 1876-4401 UR - https://doi.org/10.1016/j.artres.2013.10.196 DO - 10.1016/j.artres.2013.10.196 ID - Rudenick2013 ER -