P2.21 AGREEMENT OF CENTRAL BLOOD PRESSURE MEASURED BY APPLANATION TONOMETRY AND ECHOTRACKING
- DOI
- 10.1016/j.artres.2013.10.082How to use a DOI?
- Abstract
Background: Some studies have shown the interest of measuring central blood pressure (CBP). Usually measured non-invasively through a function transfer by applanation tonometry (Sphygmocor®) CBP can also be calculated by echotracking (Art.Lab®) at the common carotid artery assuming the fact that the distension waveform is similar to a pressure curve and thus be more precise. However, the agreement between CBP derived from these two devices remains not clear.
Objectives: To analyse the agreement between CBP and study the discrepancies between the two techniques.
Materials and methods: 74 Patients from Pharmacology Department in Paris and from Surgery, Medical, Molecular, and Critical area pathology Department at Pisa university were matched for age, sex and BP. CBP was estimated by the onboard transfer function of the Sphygmocor® derived from radial measurement. Echotracking measurements were performed on right CCA (ART.LAB) and treated to obtain CBP. Pearson’s correlation and Bland&Altman graph were performed for carotid SBP and PP, and multivariate regressions to found the determinants of discrepancies.
Results: The correlations between the carotid BP were good (SBP: r=0.79; PP: r=.77, p<0.001). We showed a small overestimation (<2 mmHg) of carotid SBP and PP by echotracking compared to applanation tonometry. Major determinants of discrepancies were center effect and age [ß=13.1(1.8), p<0.001; ß=0.1(0.01), p=0.049 respectively] for carotid PP and only center [ß=11.8(2.1), p<0.001] for carotid SBP.
Conclusion: This study shows a good agreement between the two techniques. Discrepancies could be explained by the differences between the two populations, and the Sphygmocor transfer function and calibration.
- Open Access
- This is an open access article distributed under the CC BY-NC license.
Download article (PDF)
View full text (HTML)
Cite this article
TY - JOUR AU - M.A. Alivon AU - M.L. Lindner AU - H.K. Khettab AU - E.B. Bozec AU - C.P. Palombo AU - S.L. Laurent AU - P.B. Boutouyrie PY - 2013 DA - 2013/11/11 TI - P2.21 AGREEMENT OF CENTRAL BLOOD PRESSURE MEASURED BY APPLANATION TONOMETRY AND ECHOTRACKING JO - Artery Research SP - 125 EP - 125 VL - 7 IS - 3-4 SN - 1876-4401 UR - https://doi.org/10.1016/j.artres.2013.10.082 DO - 10.1016/j.artres.2013.10.082 ID - Alivon2013 ER -