P63 Accuracy (Validation) of Central Blood Pressure Measurement Using the Sphygmocor Xcel-cuff Device

Authors
Martin Schultz1, *, Dean Picone1, Matthew Armstrong1, Andrew Black2, Nathan Dwyer2, Philip Roberts-Thomson2, James Sharman1
1Menzies Institute for Medical Research, University of Tasmania, Hobart, Australia
2Royal Hobart Hospital, Hobart, Australia
*Corresponding author. Email: martin.schultz@utas.edu.au
Corresponding Author
Martin Schultz
Available Online 17 February 2020.
DOI
10.2991/artres.k.191224.094How to use a DOI?
Abstract

Background: Numerous devices purport to measure central aortic BP as distinct from conventional brachial BP. This study aimed to determine the accuracy (validation) of the Sphygmocor Xcel-cuff device (AtCor Medical, Sydney, Australia) for measuring central BP.

Methods: 330 patients (mean age 61.3 ± 10.6 years) undergoing coronary angiography had simultaneous measurement of invasive aortic BP and non-invasive cuff-derived central BP using the Xcel device (total n = 552 individual comparisons). Methods were undertaken according to Artery Society guidelines and several calibration techniques to derive central SBP were examined.

Results: Central SBP was significantly underestimated, and with wide variability, when using the default calibration of brachial cuff SBP/DBP (−7.7 ± 11.0 mmHg). Similar wide variability was observed using other calibration methods (cuff 33% form-factor MAP/DBP, −4.4 ± 11.5 mmHg; cuff 40% form-factor MAP/DBP, 4.7 ± 11.9 mmHg; cuff oscillometric MAP/DBP, −18.2 ± 12.1 mmHg). Only calibration with invasive aortic integrated MAP/DBP resulted in a mean difference ± SD (3.3 ± 7.5 mmHg) within the minimum tolerable error of ≤5 ± ≤8 mmHg. The difference between brachial cuff SBP and invasive aortic SBP was 3.3 ± 10.7 mmHg. A subsample (n = 151) analysis to determine the accuracy of central-to-brachial SBP amplification, showed this to be over-estimated by the Xcel device (4.3 ± 9.1 mmHg, p = 0.02).

Conclusion: Irrespective of calibration technique, the Sphygmocor Xcel-cuff device does not pass the Artery Society accuracy (validation) criteria for non-invasive measurement of central BP. Further accuracy refinements of this device are required.

Copyright
© 2019 Association for Research into Arterial Structure and Physiology. Publishing services by Atlantis Press International B.V.
Open Access
This is an open access article distributed under the CC BY-NC 4.0 license (http://creativecommons.org/licenses/by-nc/4.0/).

Journal
Artery Research
Volume-Issue
25 - Supplement 1
Pages
S105 - S105
Publication Date
2020/02/17
ISSN (Online)
1876-4401
ISSN (Print)
1872-9312
DOI
10.2991/artres.k.191224.094How to use a DOI?
Copyright
© 2019 Association for Research into Arterial Structure and Physiology. Publishing services by Atlantis Press International B.V.
Open Access
This is an open access article distributed under the CC BY-NC 4.0 license (http://creativecommons.org/licenses/by-nc/4.0/).

Cite this article

TY  - JOUR
AU  - Martin Schultz
AU  - Dean Picone
AU  - Matthew Armstrong
AU  - Andrew Black
AU  - Nathan Dwyer
AU  - Philip Roberts-Thomson
AU  - James Sharman
PY  - 2020
DA  - 2020/02/17
TI  - P63 Accuracy (Validation) of Central Blood Pressure Measurement Using the Sphygmocor Xcel-cuff Device
JO  - Artery Research
SP  - S105
EP  - S105
VL  - 25
IS  - Supplement 1
SN  - 1876-4401
UR  - https://doi.org/10.2991/artres.k.191224.094
DO  - 10.2991/artres.k.191224.094
ID  - Schultz2020
ER  -