P5.14 THIGH-CUFF BASED MEASUREMENT OF AORTIC PULSE WAVE VELOCITY: INITIAL TESTING OF A NOVEL VASERA PROTOTYPE DEVICE
- DOI
- 10.1016/j.artres.2015.10.278How to use a DOI?
- Abstract
Introduction: Fully automated cuff-based devices have been developed for the assessment of arterial stiffness via pulse wave velocity (PWV) measurement, such as the VaSera device (Fukuda Denshi). To date, measurements were confined to the heart-to-ankle segment, yielding PWV and stiffness indices that are not easily inked to carotid-femoral PWV (cf-PWV), the presumed reference for measurement of aortic PWV.
Methods: We performed initial tests (N=14, 9 males, mean age 27.4±3.3, BMI 23.8±3.4) using a novel thigh-cuff prototype that can be used as a substitute for the ankle cuffs in the VaSera device. Extracted data included heart-thigh (ht-PWV) and heart-ankle (ha-PWV). cf-PWV was obtained using ultrasound (GE Vivid 7) on the right side.
Results: Measurements were successfully obtained for all subjects. cf-PWV was 5.32 ± 0.43 m/s. ha-PWV was 6.35±0.49 m/s, and was significantly higher than cf-PWV (paired t-test; P<0.001). ht-PWV, on the other hand, was 5.51±0.50 m/s and was not significantly different from cf-PWV. Bland-Altman analysis demonstrated a non-significant bias of 0.19±0.54 m/s of ht-PWV with respect to cf-PWV.
Conclusion: We conclude that, in this small-sized young and healthy population, fully automated measurement of heart-thigh PWV is straightforward and easy. Measured values were not different from carotid-femoral PWV. Further research is warranted to confirm these findings in a larger population spanning a large age range and cardiovascular risk profiles.
- Open Access
- This is an open access article distributed under the CC BY-NC license.
Download article (PDF)
View full text (HTML)
Cite this article
TY - JOUR AU - Francisco Londoño* AU - Daime Campos AU - Shigeo Horinaka AU - Julio Chirinos AU - Patrick Segers PY - 2015 DA - 2015/11/23 TI - P5.14 THIGH-CUFF BASED MEASUREMENT OF AORTIC PULSE WAVE VELOCITY: INITIAL TESTING OF A NOVEL VASERA PROTOTYPE DEVICE JO - Artery Research SP - 23 EP - 23 VL - 12 IS - C SN - 1876-4401 UR - https://doi.org/10.1016/j.artres.2015.10.278 DO - 10.1016/j.artres.2015.10.278 ID - Londoño*2015 ER -