P138 INFLUENCE OF ULTRASOUND SETTINGS ON CAROTID BIOMARKER ASSESSMENT BY B-MODE IMAGE PROCESSING

Authors
Martina Francesconi1, 2, Vincenzo Gemignani1, Rachele Gherardini2, Nicole Di Lascio1, Francesco Faita1, Lorenzo Ghiadoni2, Rosa Maria Bruno2, Elisabetta Bianchini1
1Institute of Clinical Physiology, National Research Council, Pisa, Italy
2Department of Clinical and Experimental Medicine, University of Pisa, Pisa, Italy
Available Online 4 December 2018.
DOI
10.1016/j.artres.2018.10.191How to use a DOI?
Abstract

Purpose/Background/Objectives: This study aims to investigate how image settings affect carotid Intima-Media-Thickness (IMT) and diameter estimation performed with CarotidStudio (Quipu Srl).

Methods: 14 healthy volunteers (24.4 ± 4.4 years; 8 men) were recruited. Longitudinal scans 3cm deep of common-carotid-arteries with the same region-of-interest were obtained by expert operator, starting from a “default” setup (central Time Gain Compensation-TGC; gain:50 dB; no persistence filter) and adjusting one parameter at a time (oblique TGC; gain:10 dB increments from 30 to 70 dB; persistence filter). The acquisition was performed twice and repeatability was expressed for each setup as coefficient of variation (CV). Significant changes (p < 0.05) with respect to the “default” setup were analysed by paired Student-t-test.

Results: A 70 dB gain results in overestimating IMT (from 0.46 mm ± 0.01 mm to 0.54 mm ± 0.01 mm), underestimating diameter (from 6.04 mm ± 0.04 mm to 5.70 mm ± 0.23 mm), increasing their CVs (from 2.14% ± 1.92% to 6.85% ± 5.23% for IMT, from 1.79% ± 1.22% to 4.77% ± 4.71% for diameter) and in 4/14 cases the estimation wasn’t possible. Distension CV increases also with 60 dB (from 5.71% ± 4.31% to 10.84% ± 5.91% and to 12.59% ± 7.97%). Persistence and TGC do not affect repeatability, whereas diameter (5.96 mm ± 0.03 mm) and distension (from 0.82 mm ± 0.02 mm to 0.66 mm ± 0.02 mm) values are influenced by persistence.

Conclusions: CarotidStudio precision does not show significant alterations except from 70 dB gain, which, due to the image grey-level saturation, results in poor image quality and modified values. Persistence does not affect precision but leads to reduced diameter and distension, possibly for the intrinsic behaviour of mathematical operator and image temporal filtering. Gain and persistence influence ultrasound biomarker estimation; the chosen setup should be documented and replicated in follow-up scans.

Open Access
This is an open access article distributed under the CC BY-NC license.

Journal
Artery Research
Volume-Issue
24 - C
Pages
119 - 119
Publication Date
2018/12/04
ISSN (Online)
1876-4401
ISSN (Print)
1872-9312
DOI
10.1016/j.artres.2018.10.191How to use a DOI?
Open Access
This is an open access article distributed under the CC BY-NC license.

Cite this article

TY  - JOUR
AU  - Martina Francesconi
AU  - Vincenzo Gemignani
AU  - Rachele Gherardini
AU  - Nicole Di Lascio
AU  - Francesco Faita
AU  - Lorenzo Ghiadoni
AU  - Rosa Maria Bruno
AU  - Elisabetta Bianchini
PY  - 2018
DA  - 2018/12/04
TI  - P138 INFLUENCE OF ULTRASOUND SETTINGS ON CAROTID BIOMARKER ASSESSMENT BY B-MODE IMAGE PROCESSING
JO  - Artery Research
SP  - 119
EP  - 119
VL  - 24
IS  - C
SN  - 1876-4401
UR  - https://doi.org/10.1016/j.artres.2018.10.191
DO  - 10.1016/j.artres.2018.10.191
ID  - Francesconi2018
ER  -