P.041 VALIDATION OF SPHYGMOCOR-PROCESSED AUGMENTATION INDEX USING CAROTID ARTERY DISTENSION WAVEFORM

Authors
L. Castello*2, E. Bozec1, S. Laurent1, P. Boutouyrie1
1European Hospital Georges Pompidou, Paris, France,
2University of Rome ’La Sapienza’, Rome, Italy
Available Online 13 June 2007.
DOI
10.1016/S1872-9312(07)70064-9How to use a DOI?
Open Access
This is an open access article distributed under the CC BY-NC license.

Arterial tonometry (AT), used for measurement of augmentation index (AI) at the central level, has been criticized because of artefacts induced by hand motion and deformation of the artery. We hypothesized that carotid distension waveforms, obtained with high definition echotracking device (walltrack system WTS-AI) could be used to derive augmentation index, a measure of wave reflection, and that those values were comparable with AT, either performed at carotid (CSP-AI) or on the radial artery (RSP-AI), using the generalized transfer function (GTF) of Sphygmocor system.

A group of 48 subjects with various diseases was studied. For WTS-AI, inflexion point was determined from the 0 crossing of the 3rd derivative. For CSP and RSP-AI, standard reports of Sphygmocor were used. WTS-AI could be determined with good precision from all tracings in all patients. Agreement between WTS-AI and CSP-AI was good (R2 = 0.83, RMSE = 5.8), with a 35% systematic underestimation by Sphygmocor: slope = 0.65 [0.56–0.73]). A weaker agreement between WTS-AI and RSP-AI was observed (R2 = 0.71, RMSE = 6.9), with comparable underestimation. This underestimation was not due to GTF, but to the applanation process since bypassing GTF had no influence on it and because AI, estimated with another type of tonometer, was correctly scaled with CSP-AI (slope 0.93 [0.81–1.05). CSP-AI and RSP-AI were not in good agreement (R2 = 0.66, RMSE = 10.7), but correctly scaled (slope = 0.87 [0.81–1.05].

In conclusion, wave reflections can be assessed from distension waveforms with good accuracy. Lower values for AI resulted from overestimation with applanation techniques rather than from underestimation with distension waveforms.

Journal
Artery Research
Volume-Issue
1 - S1
Pages
S37 - S37
Publication Date
2007/06/13
ISSN (Online)
1876-4401
ISSN (Print)
1872-9312
DOI
10.1016/S1872-9312(07)70064-9How to use a DOI?
Open Access
This is an open access article distributed under the CC BY-NC license.

Cite this article

TY  - JOUR
AU  - L. Castello*
AU  - E. Bozec
AU  - S. Laurent
AU  - P. Boutouyrie
PY  - 2007
DA  - 2007/06/13
TI  - P.041 VALIDATION OF SPHYGMOCOR-PROCESSED AUGMENTATION INDEX USING CAROTID ARTERY DISTENSION WAVEFORM
JO  - Artery Research
SP  - S37
EP  - S37
VL  - 1
IS  - S1
SN  - 1876-4401
UR  - https://doi.org/10.1016/S1872-9312(07)70064-9
DO  - 10.1016/S1872-9312(07)70064-9
ID  - Castello*2007
ER  -