Artery Research

Volume 1, Issue 2, September 2007, Pages 70 - 71

P.078 VALIDATION OF A NEW AUTOMATED REAL-TIME MEASURING SYSTEM FOR CAROTID INTIMA–MEDIA THICKNESS ASSESSMENT

Authors
F. Faita1, C. Giannarelli2, V. Gemignani1, E. Bianchini1, M. Demi3, L. Ghiadoni2
1CNR Institute of Clinical Physiology, Pisa, PI, Italy
2University of Pisa, Department of Intenal Medicine, Pisa, PI, Italy
3Esaote Spa, Genova, GE, Italy
Available Online 30 August 2007.
DOI
10.1016/j.artres.2007.07.012How to use a DOI?
Open Access
This is an open access article distributed under the CC BY-NC license.

Increased carotid intima-media thickness (C-IMT) is a non-invasive marker of early arterial alterations which is associated with increased cardiovascular risk. The aim of this study is to validate a new real-time automatic system to measure C-IMT. Results are compared with the manual measurements, which are still largely accepted as the gold standard.

Ultrasound scans of the distal common carotid artery (right/left) were obtained from 120 patients with cardiovascular risk factors and 30 healthy controls. The C-IMT was measured on the far wall, 1 cm above the bifurcation. The dataset was analysed by two operators both automatically and manually. The first operator repeated the analysis twice.

The agreement between automatic and manual measurements was evaluated by Bland-Altman plots: a bias of −0.020mm and an interval of agreement of 0.027mm were obtained. Intra-observer variability was computed on the repeated measurements of the first operator. Bias was not significantly different from zero for both manual and automatic measurements, whereas the interval of agreement was 0.077mm in manual analysis and 0.012mm in automatic analysis. Coefficients of variation of 2.8% and 0.4% were obtained, respectively. Inter-observer variability showed a little bias (−0.032mm) only for manual analysis, whereas the interval of agreement was 0.075mm in manual analysis and 0.021mm in automatic analysis with coefficients of variation of 4.5% and 0.6%, respectively.

In conclusion, the new real-time automatic system represents a more feasible and reproducible method than the manual approach when used to estimate C-IMT in clinical studies and practice.

Journal
Artery Research
Volume-Issue
1 - 2
Pages
70 - 71
Publication Date
2007/08/30
ISSN (Online)
1876-4401
ISSN (Print)
1872-9312
DOI
10.1016/j.artres.2007.07.012How to use a DOI?
Open Access
This is an open access article distributed under the CC BY-NC license.

Cite this article

TY  - JOUR
AU  - F. Faita
AU  - C. Giannarelli
AU  - V. Gemignani
AU  - E. Bianchini
AU  - M. Demi
AU  - L. Ghiadoni
PY  - 2007
DA  - 2007/08/30
TI  - P.078 VALIDATION OF A NEW AUTOMATED REAL-TIME MEASURING SYSTEM FOR CAROTID INTIMA–MEDIA THICKNESS ASSESSMENT
JO  - Artery Research
SP  - 70
EP  - 71
VL  - 1
IS  - 2
SN  - 1876-4401
UR  - https://doi.org/10.1016/j.artres.2007.07.012
DO  - 10.1016/j.artres.2007.07.012
ID  - Faita2007
ER  -