P2.14 INFLUENCE OF CALIBRATION OF PERIPHERAL PRESSURE ON THE ESTIMATION OF CENTRAL SYSTOLIC BLOOD PRESSURE

Authors
A. Guilcher, B. Clapp, S. Brett, P. Chowienczyk
King’s College London, School of Medicine, Cardiovascular Division, St Thomas’ Hospital, London, United Kingdom
Available Online 15 September 2008.
DOI
10.1016/j.artres.2008.08.380How to use a DOI?
Open Access
This is an open access article distributed under the CC BY-NC license.

Objective: We compared the accuracy with which central systolic aortic pressure (cSBP) can be estimated from the late systolic shoulder of the digital artery pressure waveform (SBP2) when the digital waveform is calibrated using a validated non-invasive oscillometric device and when it is calibrated using mean (MAP) and diastolic (DBP) central blood pressure measurement.

Design: Subjects (n=25) were studied at the time of cardiac catheterisation for diagnostic angiography and/or angioplasty. The study was approved by the local research ethics committee and all subjects gave written informed consent.

Methods: cSBP was measured with a Millar SPC-454D (Millar instruments Houston, Texas) catheter with the tip of the catheter in the proximal aortic root. Peripheral pressure waveforms were acquired from the digital artery using a Finometer (Finapres medical systems, Netherlands) and were calibrated from aortic mean and diastolic pressures and from systolic and diastolic pressures measured using an Omron 750IT (Omron Healthcare). Measurements of digital artery and aortic pressures were obtained at baseline and after nitroglycerin (NTG, sublingual spray 500 μg).

Results: The overall mean (±SD) difference between cSBP and SBP2 calibrated using the Omron 750IT was 1.4±11.8 mmHg. When SBP2 was calibrated from aortic MAP and DBP the difference between cSBP and SBP2 was −1.1±5.6 mmHg.

Conclusion: These results suggest non-invasive calibration does not produce a major systematic error in estimation of cSBP from SBP2 but does introduce greater variability when compared to invasive calibration.

Journal
Artery Research
Volume-Issue
2 - 3
Pages
109 - 110
Publication Date
2008/09/15
ISSN (Online)
1876-4401
ISSN (Print)
1872-9312
DOI
10.1016/j.artres.2008.08.380How to use a DOI?
Open Access
This is an open access article distributed under the CC BY-NC license.

Cite this article

TY  - JOUR
AU  - A. Guilcher
AU  - B. Clapp
AU  - S. Brett
AU  - P. Chowienczyk
PY  - 2008
DA  - 2008/09/15
TI  - P2.14 INFLUENCE OF CALIBRATION OF PERIPHERAL PRESSURE ON THE ESTIMATION OF CENTRAL SYSTOLIC BLOOD PRESSURE
JO  - Artery Research
SP  - 109
EP  - 110
VL  - 2
IS  - 3
SN  - 1876-4401
UR  - https://doi.org/10.1016/j.artres.2008.08.380
DO  - 10.1016/j.artres.2008.08.380
ID  - Guilcher2008
ER  -